Abstract
The results of 115 primary total hip arthroplasties (97 patients) performed using the Charnley system and contemporary cementing techniques were reviewed. Fifty-five all-polyethylene cups and 60 metal-backed cups were used. The mean follow-up period was 84 months. No statistically significant difference was noted between groups for radiographic loosening and no cup has been revised to date. Survival analysis did not reveal a statistically significant difference between groups. Although all-polyethylene cups had significantly greater polyethylene thickness, no difference was noted for rate of wear. Increased loosening of the metal-backed cups was not observed. Polyethylene thickness is increased by the use of 22-mm heads and this may have accounted for the lack of difference in loosening rates of the all-polyethylene and metal-backed cups.
Original language | English (US) |
---|---|
Pages (from-to) | S1-S7 |
Journal | The Journal of Arthroplasty |
Volume | 10 |
Issue number | SUPPL. 1 |
DOIs | |
State | Published - Nov 1995 |
Keywords
- all-polyethylene
- cemented acetabulae
- hip arthroplasty
- hip joint
- joint arthroplasty
- metal-backed