Use of a pressure sensing sheath: Comparison with standard means of blood pressure monitoring in catheterization procedures

Phillip D. Purdy, Charles South, Richard P. Klucznik, Kenneth C. Liu, Robin L. Novakovic, Ajit S. Puri, G. Lee Pride, Beverly Aagaard-Kienitz, Abishek Ray, Alan C. Elliott

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Purpose Monitoring of blood pressure (BP) during procedures is variable, depending on multiple factors. Common methods include sphygmomanometer (BP cuff), separate radial artery catheterization, and side port monitoring of an indwelling sheath. Each means of monitoring has disadvantages, including time consumption, added risk, and signal dampening due to multiple factors. We sought an alternative approach to monitoring during procedures in the catheterization laboratory. Methods A new technology involving a 330 μm fiberoptic sensor embedded in the wall of a sheath structure was tested against both radial artery catheter and sphygmomanometer readings obtained simultaneous with readings recorded from the pressure sensing system (PSS). Correlations and Bland-Altman analysis were used to determine whether use of the PSS could substitute for these standard techniques. Results The results indicated highly significant correlations in systolic, diastolic, and mean arterial pressures (MAP) when compared against radial artery catheterization (p<0.0001), and MAP means differed by <4%. Bland-Altman analysis of the data suggested that the sheath measurements can replace a separate radial artery catheter. While less striking, significant correlations were seen when PSS readings were compared against BP cuff readings. Conclusions The PSS has competitive functionality to that seen with a dedicated radial artery catheter for BP monitoring and is available immediately on sheath insertion without the added risk of radial catheterization. The sensor is structurally separated from the primary sheath lumen and readings are unaffected by device introduction through the primary lumen. Time delays and potential complications from radial artery catheterization are avoided.

Original languageEnglish (US)
Pages (from-to)766-771
Number of pages6
JournalJournal of NeuroInterventional Surgery
Volume9
Issue number8
DOIs
StatePublished - Aug 1 2017

Keywords

  • Angiography
  • Blood Pressure
  • Catheter
  • Device

ASJC Scopus subject areas

  • Surgery
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'Use of a pressure sensing sheath: Comparison with standard means of blood pressure monitoring in catheterization procedures'. Together they form a unique fingerprint.

Cite this