The relationship between ventricular fluid pressure and body position in normal subjects and subjects with shunts: A telemetric study

P. H. Chapman, E. R. Cosman, M. A. Arnold, H. D. Portnoy, F. Sklar

Research output: Contribution to journalArticlepeer-review

180 Scopus citations

Abstract

Using a chronically implanted telemetric pressure sensor, we have determined the quantitative relationship between changes in body position and ventricular fluid pressure in normal subjects and subjects with shunts. The method allows accurate, reliable measurement of negative as well as positive pressures. We describe the derangement of postural intraventricular pressure regulation caused by placement of a shunt, as well as the influence of various shunt systems and the antisiphon device on this problem. Ventriculoatrial, venticuloperitoneal, and ventriculopleural shunt systems and the antisiphon device on this problem. Ventriculoatrial, ventriculoperitoneal, and ventriculopleural shunts all caused similar severely abnormal postural intracranial pressure relationships. The antisiphon device was generally effective in restoring normal pressures in patients in the upright position. We discuss the implications of our findings for understanding the mechanisms of postural intracranial pressure regulation in patients without hydrocephalus.

Original languageEnglish (US)
Pages (from-to)181-189
Number of pages9
JournalNeurosurgery
Volume26
Issue number2
DOIs
StatePublished - 1990

Keywords

  • antisiphon device
  • cerebrospinal fluid shunts
  • hydrocephalus
  • intracranial pressure
  • telemetry

ASJC Scopus subject areas

  • Surgery
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'The relationship between ventricular fluid pressure and body position in normal subjects and subjects with shunts: A telemetric study'. Together they form a unique fingerprint.

Cite this