T1 - Proactive esophageal cooling protects against thermal insults during high-power short-duration radiofrequency cardiac ablation
AU - Mercado Montoya, Marcela
AU - Gomez Bustamante, Tatiana
AU - Berjano, Enrique
AU - Mickelsen, Steven R.
AU - Daniels, James D.
AU - Hernandez Arango, Pablo
AU - Schieber, Jay
AU - Kulstad, Erik
N1 - Funding Information:
Research reported in this publication was supported by the National Heart, Lung, And Blood Institute of the National Institutes of Health under Award Number R44HL158375 (the content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health) and by the Spanish Ministerio de Ciencia, Innovación y Universidades/Agencia Estatal de Investigación (MCIN/AEI/10.13039/501100011033 under grant RTI2018-094357-B-C21).
Publisher Copyright:
© 2022 The Author(s). Published with license by Taylor & Francis Group, LLC.
PY - 2022
Y1 - 2022
N2 - Background: Proactive cooling with a novel cooling device has been shown to reduce endoscopically identified thermal injury during radiofrequency (RF) ablation for the treatment of atrial fibrillation using medium power settings. We aimed to evaluate the effects of proactive cooling during high-power short-duration (HPSD) ablation. Methods: A computer model accounting for the left atrium (1.5 mm thickness) and esophagus including the active cooling device was created. We used the Arrhenius equation to estimate the esophageal thermal damage during 50 W/ 10 s and 90 W/ 4 s RF ablations. Results: With proactive esophageal cooling in place, temperatures in the esophageal tissue were significantly reduced from control conditions without cooling, and the resulting percentage of damage to the esophageal wall was reduced around 50%, restricting damage to the epi-esophageal region and consequently sparing the remainder of the esophageal tissue, including the mucosal surface. Lesions in the atrial wall remained transmural despite cooling, and maximum width barely changed (<0.8 mm). Conclusions: Proactive esophageal cooling significantly reduces temperatures and the resulting fraction of damage in the esophagus during HPSD ablation. These findings offer a mechanistic rationale explaining the high degree of safety encountered to date using proactive esophageal cooling, and further underscore the fact that temperature monitoring is inadequate to avoid thermal damage to the esophagus.
AB - Background: Proactive cooling with a novel cooling device has been shown to reduce endoscopically identified thermal injury during radiofrequency (RF) ablation for the treatment of atrial fibrillation using medium power settings. We aimed to evaluate the effects of proactive cooling during high-power short-duration (HPSD) ablation. Methods: A computer model accounting for the left atrium (1.5 mm thickness) and esophagus including the active cooling device was created. We used the Arrhenius equation to estimate the esophageal thermal damage during 50 W/ 10 s and 90 W/ 4 s RF ablations. Results: With proactive esophageal cooling in place, temperatures in the esophageal tissue were significantly reduced from control conditions without cooling, and the resulting percentage of damage to the esophageal wall was reduced around 50%, restricting damage to the epi-esophageal region and consequently sparing the remainder of the esophageal tissue, including the mucosal surface. Lesions in the atrial wall remained transmural despite cooling, and maximum width barely changed (<0.8 mm). Conclusions: Proactive esophageal cooling significantly reduces temperatures and the resulting fraction of damage in the esophagus during HPSD ablation. These findings offer a mechanistic rationale explaining the high degree of safety encountered to date using proactive esophageal cooling, and further underscore the fact that temperature monitoring is inadequate to avoid thermal damage to the esophagus.
KW - Atrial fibrillation
KW - atrioesophageal fistula
KW - esophageal cooling
KW - mathematical modeling
KW - radiofrequency ablation
UR - http://www.scopus.com/inward/record.url?scp=85137880265&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85137880265&partnerID=8YFLogxK
U2 - 10.1080/02656736.2022.2121860
DO - 10.1080/02656736.2022.2121860
M3 - Article
C2 - 36104029
AN - SCOPUS:85137880265
SN - 0265-6736
VL - 39
SP - 1202
EP - 1212
JO - International Journal of Hyperthermia
JF - International Journal of Hyperthermia
IS - 1
ER -