Parameters for direct cortical electrical stimulation in the human: histopathologic confirmation

Barry Gordon, Ronald P. Lesser, Naomi E. Rance, John Hart, Robert Webber, Sumio Uematsu, Robert S. Fisher

Research output: Contribution to journalArticlepeer-review

192 Scopus citations

Abstract

Safe parameters for electrical cortical stimulation in humans are difficult to estimate from the animal experimental literature. We therefore examined the light microscopic histology at a total of 11 sites of direct subdural electrical stimulation, taken as part of anterior temporal lobectomies in 3 patients. Stimulations had been done through 3.175 mm diameter electrodes, with 0.3 msec square wave pulses of alternating polarity at 50 pulses/sec. In 2 patients, one site each had been used as a common reference for stimulation, receiving over 251 stimulation trials, most of 2-5 sec duration, at currents of 12.5-15.0 mA, 1 day prior to resection. The maximum charge per phase was 4.0-4.4 μC; the maximum charge density was 52-57 μC per geometric cm2 per pulse at the electrode surfaces. Comparison of hematoxylin and eosin, periodic acid-Schiff, and cresyl violet-stained material from the electrode sites with that from other regions did not show any histologic abnormalities attributable to the electrical stimulation. The relatively brief and intermittent periods utilized for human stimulation testing do not appear to cause strcutural damage at the light microscopic level at charge densities that exceed the threshold for damage established in animal studies with more continuous, chronic stimulation schedules.

Original languageEnglish (US)
Pages (from-to)371-377
Number of pages7
JournalElectroencephalography and Clinical Neurophysiology
Volume75
Issue number5
DOIs
StatePublished - May 1990

Keywords

  • (Human)
  • Electrical cortical stimulation
  • Histopathology
  • Safety
  • Subdural arrays

ASJC Scopus subject areas

  • General Neuroscience
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'Parameters for direct cortical electrical stimulation in the human: histopathologic confirmation'. Together they form a unique fingerprint.

Cite this