J-Difference editing (MEGA) of lactate in the human brain at 3T

Ryan K. Robison, Justin R. Haynes, Sandeep K. Ganji, Charles P. Nockowski, Zoltan Kovacs, Wellington Pham, Victoria L. Morgan, Seth A. Smith, Reid C. Thompson, Reed A. Omary, John C. Gore, Changho Choi

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

Purpose: The need to detect and quantify brain lactate accurately by MRS has stimulated the development of editing sequences based on J coupling effects. In J-difference editing of lactate, threonine can be co-edited and it contaminates lactate estimates due to the spectral proximity of the coupling partners of their methyl protons. We therefore implemented narrow-band editing 180° pulses (E180) in MEGA-PRESS acquisitions to resolve separately the 1.3-ppm resonances of lactate and threonine. Methods: Two 45.3-ms rectangular E180 pulses, which had negligible effects 0.15-ppm away from the carrier frequency, were implemented in a MEGA-PRESS sequence with TE 139 ms. Three acquisitions were designed to selectively edit lactate and threonine, in which the E180 pulses were tuned to 4.1 ppm, 4.25 ppm, and a frequency far off resonance. Editing performance was validated with numerical analyses and acquisitions from phantoms. The narrow-band E180 MEGA and another MEGA-PRESS sequence with broad-band E180 pulses were evaluated in six healthy subjects. Results: The 45.3-ms E180 MEGA offered a difference-edited lactate signal with lower intensity and reduced contamination from threonine compared to the broad-band E180 MEGA. The 45.3 ms E180 pulse had MEGA editing effects over a frequency range larger than seen in the singlet-resonance inversion profile. Lactate and threonine in healthy brain were both estimated to be 0.4 ± 0.1 mM, with reference to N-acetylaspartate at 12 mM. Conclusion: Narrow-band E180 MEGA editing minimizes threonine contamination of lactate spectra and may improve the ability to detect modest changes in lactate levels.

Original languageEnglish (US)
Pages (from-to)852-862
Number of pages11
JournalMagnetic resonance in medicine
Volume90
Issue number3
DOIs
StatePublished - Sep 2023

Keywords

  • 3T
  • H MRS
  • J-difference editing (MEGA)
  • lactate
  • threonine

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'J-Difference editing (MEGA) of lactate in the human brain at 3T'. Together they form a unique fingerprint.

Cite this