Influence of tumor shape and location in eye plaque brachytherapy dosimetry

Matthew T. Studenski, Nirav V. Patel, Arnold Markoe, J. William Harbour, Stuart E. Samuels

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Purpose: A common treatment planning technique for eye plaque brachytherapy is to model the tumor as an ellipse. For posterior tumors near the optic disc and fovea, this approach may lead to overlap between tumor and the organ at risk (OAR). We hypothesized that a superior plan can be generated by modeling the actual tumor shape. Materials and Methods: Forty eye plaque patients with tumors <1 cm from the optic disc and fovea were selected. Two treatment plans were generated for each patient: an elliptical tumor model plan and a true tumor model plan. Dosimetric data were collected for each plan, and Wilcoxon signed-rank tests were used to asses any statistically significant differences. Results: Equivalent tumor coverage was confirmed between the elliptical and true tumor plans for all patients. Qualitative analysis showed greater dosimetric differences between plans as the distance from the OARs increased from 0 to 2 mm but the largest differences were observed between 2 and 4 mm. Minimal differences between models were seen beyond 4 mm. Statistically significant dosimetric improvements were found for tumors <4 mm from the fovea and <2 mm from the optic disc. Conclusions: Intuitively, accurate modeling of the tumor accounting for irregularities in the shape should result in a more conformal plan and an overall reduction in OAR dose. However, this technique is only beneficial for tumors that are within 4 mm of the fovea or optic disc. An elliptical tumor model allows for an acceptable plan unless the tumor is located posteriorly and has an irregular shape.

Original languageEnglish (US)
Pages (from-to)249-254
Number of pages6
JournalBrachytherapy
Volume19
Issue number2
DOIs
StatePublished - Mar 1 2020
Externally publishedYes

Keywords

  • Brachytherapy
  • Dosimetry
  • Eye plaque

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Influence of tumor shape and location in eye plaque brachytherapy dosimetry'. Together they form a unique fingerprint.

Cite this