T1 - Diffusion tensor imaging of the median nerve at 3.0 T using different MR scanners
T2 - Agreement of FA and ADC measurements
AU - Guggenberger, Roman
AU - Nanz, Daniel
AU - Bussmann, Lorenz
AU - Chhabra, Avneesh
AU - Fischer, Michael A.
AU - Hodler, Jürg
AU - Pfirrmann, Christian W A
AU - Andreisek, Gustav
PY - 2013/10
Y1 - 2013/10
N2 - Objective: To assess the agreement of fractional anisotropy (FA) and apparent diffusion coefficient (ADC) values of the median nerve on 3.0 T MR scanners from different vendors. Materials and methods: IRB approved study including 16 healthy volunteers (9 women; mean age 30.6 ± 5.3 years). Diffusion tensor imaging (DTI) of the dominant wrist was performed on three 3.0 T MR scanners (GE, Siemens, Philips) using similar imaging protocols and vendor-proprietary hard- and software. Intra-, inter-reader and inter-vendor agreements were assessed. Results: ICCs for intra-/inter-reader agreements ranged from 0.843-0.970/0.846-0.956 for FA, and 0.840-0.940/0.726-0.929 for ADC, respectively. ANOVA analysis identified significant differences for FA/ADC measurements among vendors (p < 0.001/p < 0.01, respectively). Overall mean values for FA were 0.63 (SD ± 0.1) and 0.999 × 10 -3mm2/s (SD ± 0.134 × 10-3) for ADC. A significant negative measurement bias was found for FA values from the GE scanner (-0.05 and -0.07) and for ADC values from the Siemens scanner (-0.053 and -0.063 × 10-3mm2/s) as compared to the remainder vendors Conclusion: FA and ADC values of the median nerve obtained on different 3.0 T MR scanners differ significantly, but are in comparison to the standard deviation of absolute values small enough to not have an impact on larger group studies or when substantial diffusion changes can be expected. However, caution is warranted in an individual patient when interpreting diffusion values from different scanner acquisitions.
AB - Objective: To assess the agreement of fractional anisotropy (FA) and apparent diffusion coefficient (ADC) values of the median nerve on 3.0 T MR scanners from different vendors. Materials and methods: IRB approved study including 16 healthy volunteers (9 women; mean age 30.6 ± 5.3 years). Diffusion tensor imaging (DTI) of the dominant wrist was performed on three 3.0 T MR scanners (GE, Siemens, Philips) using similar imaging protocols and vendor-proprietary hard- and software. Intra-, inter-reader and inter-vendor agreements were assessed. Results: ICCs for intra-/inter-reader agreements ranged from 0.843-0.970/0.846-0.956 for FA, and 0.840-0.940/0.726-0.929 for ADC, respectively. ANOVA analysis identified significant differences for FA/ADC measurements among vendors (p < 0.001/p < 0.01, respectively). Overall mean values for FA were 0.63 (SD ± 0.1) and 0.999 × 10 -3mm2/s (SD ± 0.134 × 10-3) for ADC. A significant negative measurement bias was found for FA values from the GE scanner (-0.05 and -0.07) and for ADC values from the Siemens scanner (-0.053 and -0.063 × 10-3mm2/s) as compared to the remainder vendors Conclusion: FA and ADC values of the median nerve obtained on different 3.0 T MR scanners differ significantly, but are in comparison to the standard deviation of absolute values small enough to not have an impact on larger group studies or when substantial diffusion changes can be expected. However, caution is warranted in an individual patient when interpreting diffusion values from different scanner acquisitions.
KW - Diffusion tensor imaging
KW - Magnetic resonance imaging
KW - Median nerve
KW - Peripheral neuropathy
UR - http://www.scopus.com/inward/record.url?scp=84893700145&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84893700145&partnerID=8YFLogxK
U2 - 10.1016/j.ejrad.2013.05.011
DO - 10.1016/j.ejrad.2013.05.011
M3 - Article
C2 - 23768611
AN - SCOPUS:84893700145
SN - 0720-048X
VL - 82
SP - e590-e596
JO - European Journal of Radiology
JF - European Journal of Radiology
IS - 10
ER -