Abstract
Purpose: This paper focused on the design and optimization of ridge filter–based intensity-modulated proton therapy (IMPT), and its potential applications for FLASH. Differing from the standard pencil beam scanning (PBS) mode, no energy/layer switching is required and total treatment time can be shortened. Methods: Unique dose-influence matrices were generated as a proton beam traverses through slabs of different thicknesses (i.e., modulation by different layers). To establish the references for comparison, conventional IMPT plans (single field) were created using a large-scale nonlinear solver. The spot weights from the reference IMPT plans were used as inputs for optimizing the design of ridge filters. Two designs were evaluated: model A (static) and model B (dynamic). The ridge filter designs were first verified (by GEANT4 simulation) in a water phantom and then in an H&N case. A direct comparison was made between the GEANT4 simulation results of two models and their respective references, with regard to plan quality, dose-averaged dose rate, and total treatment time. Results: In both the water phantom and the H&N case, two models are able to modulate dose distributions with high conformity, showing no significant difference relative to the reference plans. Dose rate–volume histograms suggest that in order to achieve a dose rate of 40 Gy/s over 90% PTV, the beam intensity needs to be 2.5 × 1011 protons/s for both models. For a fraction dose of 10 Gy, the total treatment time (including both irradiation time and dead time) can be shortened by a factor of 4.9 (model A) and 6.5 (model B), relative to the reference plans. Conclusion: Two proposed designs (both static and dynamic) can be used for PBS–IMPT requiring no layer switching. They are promising candidates for FLASH-IMPT capable of reducing treatment time and achieving high dose rates while maintaining dose conformity simultaneously.
Original language | English (US) |
---|---|
Pages (from-to) | 5387-5399 |
Number of pages | 13 |
Journal | Medical physics |
Volume | 49 |
Issue number | 8 |
DOIs | |
State | Published - Aug 2022 |
Externally published | Yes |
Keywords
- FLASH
- IMPT
- dose rate
- proton therapy
- ridge filter
ASJC Scopus subject areas
- Biophysics
- Radiology Nuclear Medicine and imaging