TY - JOUR
T1 - Comparison of Spectra Optia and COBE Spectra apheresis systems' performances for red blood cell exchange procedures
AU - Kim, Jaehyup
AU - Joseph, Ranjit
AU - Matevosyan, Karen
AU - Sarode, Ravi
PY - 2016/12/1
Y1 - 2016/12/1
N2 - Background Spectra Optia (Terumo BCT, Lakewood, CO) was FDA approved for red blood cell exchange (RBCx) procedures in January 2014 and is expected to replace COBE spectra (Terumo BCT) very soon in the USA. The performance characteristics of these devices for Isovolemic Hemodilution (IHD-RBCx) procedure were compared in this study. Methods A total of 114 IHD-RBCx procedures from 19 patients were analyzed. For every patient, three procedures on each device with similar pre-procedure hematocrits were compared. Pre and post procedure laboratory parameters compared were hemoglobin S (HbS), hematocrits (Hct), platelet counts and fraction of cells remaining (FCR). Statistical analysis was performed using t-test adjusted by the Holm-Bonferroni method to reduce family-wise error rate. Results There were no significant differences between these two devices in regards to HbS, Hct, FCR and platelet counts (p = > 0.05). However, rinseback volume (124.2 ± 8.9 ml) and normal saline replacement volume during IHD phase (296.1 ± 97.2 ml) were lower in Spectra Optia as compared to COBE Spectra (337 ± 33.8 ml and 326.6 ± 105.2 ml, p value <0.001 and 0.030 respectively). Spectra Optia had a longer run time (107.1 ± 15.9 min vs 123.8 ± 19.6 min, p value <0.001) overall. Conclusions Performance characteristics of Spectra Optia for HbS, Hct and FCR were similar to COBE Spectra for IHD-RBCx. IHD-RBCx procedure on Optia required less normal saline replacement volume and rinse back volume but with overall longer procedure run time.
AB - Background Spectra Optia (Terumo BCT, Lakewood, CO) was FDA approved for red blood cell exchange (RBCx) procedures in January 2014 and is expected to replace COBE spectra (Terumo BCT) very soon in the USA. The performance characteristics of these devices for Isovolemic Hemodilution (IHD-RBCx) procedure were compared in this study. Methods A total of 114 IHD-RBCx procedures from 19 patients were analyzed. For every patient, three procedures on each device with similar pre-procedure hematocrits were compared. Pre and post procedure laboratory parameters compared were hemoglobin S (HbS), hematocrits (Hct), platelet counts and fraction of cells remaining (FCR). Statistical analysis was performed using t-test adjusted by the Holm-Bonferroni method to reduce family-wise error rate. Results There were no significant differences between these two devices in regards to HbS, Hct, FCR and platelet counts (p = > 0.05). However, rinseback volume (124.2 ± 8.9 ml) and normal saline replacement volume during IHD phase (296.1 ± 97.2 ml) were lower in Spectra Optia as compared to COBE Spectra (337 ± 33.8 ml and 326.6 ± 105.2 ml, p value <0.001 and 0.030 respectively). Spectra Optia had a longer run time (107.1 ± 15.9 min vs 123.8 ± 19.6 min, p value <0.001) overall. Conclusions Performance characteristics of Spectra Optia for HbS, Hct and FCR were similar to COBE Spectra for IHD-RBCx. IHD-RBCx procedure on Optia required less normal saline replacement volume and rinse back volume but with overall longer procedure run time.
KW - COBE Spectra
KW - Red blood cell exchange
KW - Sickle cell anemia
KW - Spectra Optia
UR - http://www.scopus.com/inward/record.url?scp=85003914641&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85003914641&partnerID=8YFLogxK
U2 - 10.1016/j.transci.2016.10.004
DO - 10.1016/j.transci.2016.10.004
M3 - Article
C2 - 27789125
AN - SCOPUS:85003914641
SN - 1473-0502
VL - 55
SP - 368
EP - 370
JO - Transfusion and Apheresis Science
JF - Transfusion and Apheresis Science
IS - 3
ER -