Arterial input function calculation in dynamic contrast-enhanced MRI: An in vivo validation study using co-registered contrast-enhanced ultrasound imaging

Hatef Mehrabian, Chaitanya Chandrana, Ian Pang, Rajiv Chopra, Anne L. Martel

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

Objectives Developing a method of separating intravascular contrast agent concentration to measure the arterial input function (AIF) in dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) of tumours, and validating its performance in phantom and in vivo experiments. Methods A tissue-mimicking phantom was constructed to model leaky tumour vasculature and DCE-MR images of this phantom were acquired. An in vivo study was performed using tumour-bearing rabbits. Co-registered DCE-MRI and contrast-enhanced ultrasound (CEUS) images were acquired. An independent component analysis (ICA)-based method was developed to separate the intravascular component from DCEMRI. Results were validated by comparing the time-intensity curves with the actual phantom and in vivo curves. Results Phantom study: the AIF extracted using ICA correlated well with the true intravascular curve. In vivo study: the AIFs extracted from DCE-MRI using ICA were very close to the true AIF. Intravascular component images were very similar to the CEUS images. The contrast onset times and initial wash-in slope of the ICA-derived AIF showed good agreement with the CEUS curves. Conclusion ICA has the potential to separate the intravascular component from DCE-MRI. This could eliminate the requirement for contrast medium uptake measurements in a major artery and potentially result in more accurate pharmacokinetic parameters. Key Points ̇ Tumour response to therapy can be inferred from pharmacokinetic parameters. ̇ Arterial input function (AIF) is required for pharmacokinetic modelling of tumours. ̇ Independent component analysis has the potential to measure AIF inside the tumour. ̇ AIF measurement is validated using contrast enhanced ultrasound and phantoms.

Original languageEnglish (US)
Pages (from-to)1735-1747
Number of pages13
JournalEuropean Radiology
Volume22
Issue number8
DOIs
StatePublished - Aug 2012

Keywords

  • Arterial input function (AIF)
  • Contrast-enhanced ultrasound (CEUS)
  • Dynamic contrast-enhanced MRI (DCE-MRI)
  • Independent component analysis (ICA)
  • Pharmacokinetics