T1 - Accuracy evaluation of using MV EPID and/or integrated kV imaging system to locate 3D positions of internal fiducial markers using a Calypso system
AU - Mao, Weihua
AU - Foster, Ryan
AU - Solberg, Timothy
N1 - Copyright:
Copyright 2021 Elsevier B.V., All rights reserved.
PY - 2009
Y1 - 2009
N2 - Introduction: In order to use MV EPID and/or integrated kV imaging system to obtained 3D positions of implanted fiducial markers for the purpose of patient setup and/or image-guided radiation therapy in real-time, it is essential to investigate the accuracy of this method. Materials and Methods: A Calypso QA phantom with three Beacons was imaged by both MV and kV beams at a series of gantry angles on a Trilogy (Varian Medical Systems, Palo Alto, California, USA) while Beacon positions were simultaneously tracked using a Calypso 4D localization system (Calypso Medical Technologies, Inc., Seattle, WA, USA). Beacons were automatically detected on kV and MV projection images using in-house software, their 3D positions were calculated from MV-MV, kV-kV, or MV-kV pairs with a 60° or larger gantry angle separation between each pair, and compared with Calypso results. A special QA procedure was developed to assess geometric accuracy of MV and kV imaging systems on the Trilogy, and geometric parameter deviations, particularly detector center offsets, were applied to correct the 3D position calculations. Results: A phantom was placed at known positions based on the detected Calypso Beacon centroids. Both MV and kV images were acquired at 12 gantry angles with an increment of 30°. 2D and 3D positions were calculated. The different calculation modalities resulted in noticeable discrepancies from the Calypso 3D positions. The application of a geometric correction is capable of reducing these deviations, with the measured net shifts matching calypso results within 0.1 mm. Conclusion: Geometric correction after QA of imaging geometry helps improve detection accuracy. Implanted fiducial markers can be detected by multiple projections of either MV or kV x-ray beams with an accuracy of approximately 1 mm.
AB - Introduction: In order to use MV EPID and/or integrated kV imaging system to obtained 3D positions of implanted fiducial markers for the purpose of patient setup and/or image-guided radiation therapy in real-time, it is essential to investigate the accuracy of this method. Materials and Methods: A Calypso QA phantom with three Beacons was imaged by both MV and kV beams at a series of gantry angles on a Trilogy (Varian Medical Systems, Palo Alto, California, USA) while Beacon positions were simultaneously tracked using a Calypso 4D localization system (Calypso Medical Technologies, Inc., Seattle, WA, USA). Beacons were automatically detected on kV and MV projection images using in-house software, their 3D positions were calculated from MV-MV, kV-kV, or MV-kV pairs with a 60° or larger gantry angle separation between each pair, and compared with Calypso results. A special QA procedure was developed to assess geometric accuracy of MV and kV imaging systems on the Trilogy, and geometric parameter deviations, particularly detector center offsets, were applied to correct the 3D position calculations. Results: A phantom was placed at known positions based on the detected Calypso Beacon centroids. Both MV and kV images were acquired at 12 gantry angles with an increment of 30°. 2D and 3D positions were calculated. The different calculation modalities resulted in noticeable discrepancies from the Calypso 3D positions. The application of a geometric correction is capable of reducing these deviations, with the measured net shifts matching calypso results within 0.1 mm. Conclusion: Geometric correction after QA of imaging geometry helps improve detection accuracy. Implanted fiducial markers can be detected by multiple projections of either MV or kV x-ray beams with an accuracy of approximately 1 mm.
KW - Fiducial marker
KW - Geometric QA
KW - IGRT
KW - KV imaging
KW - MV imaging
UR - http://www.scopus.com/inward/record.url?scp=77950384925&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77950384925&partnerID=8YFLogxK
U2 - 10.1007/978-3-642-03474-9_209
DO - 10.1007/978-3-642-03474-9_209
M3 - Conference contribution
AN - SCOPUS:77950384925
SN - 9783642034725
T3 - IFMBE Proceedings
SP - 740
EP - 743
BT - World Congress on Medical Physics and Biomedical Engineering
PB - Springer Verlag
T2 - World Congress on Medical Physics and Biomedical Engineering: Radiation Oncology
Y2 - 7 September 2009 through 12 September 2009
ER -