TY - JOUR
T1 - Severe symptomatic nickel allergy following stent graft implantation requiring excision and external iliac artery reconstruction
AU - Patel, Rhusheet
AU - Moore, Wesley
AU - Jimenez, Juan Carlos
N1 - Funding Information:
The editors and reviewers of this article have no relevant financial relationships to disclose per the Journal policy that requires reviewers to decline review of any manuscript for which they may have a conflict of interest.
Publisher Copyright:
© 2022 The Author(s)
PY - 2022/12
Y1 - 2022/12
N2 - Although nickel allergy is a common cause of contact dermatitis, systemic reactions to nitinol stents are rare. A 61-year-old woman had presented with a nonhealing toe wound. Angiography revealed an external iliac artery stenosis, which was treated with a nitinol stent graft. However, she developed severe truncal pruritus, and within 3 months, her external iliac stent graft had thrombosed. Allergy testing revealed nickel sensitivity. After medical therapy had failed, stent graft removal was performed, resulting in complete resolution of her symptoms. The present case demonstrates a rare allergic reaction to the nitinol in commercially available stent grafts. Pruritus and rash are rare reactions to stenting; however, a nitinol allergy should be considered for patients with no other identifiable primary source.
AB - Although nickel allergy is a common cause of contact dermatitis, systemic reactions to nitinol stents are rare. A 61-year-old woman had presented with a nonhealing toe wound. Angiography revealed an external iliac artery stenosis, which was treated with a nitinol stent graft. However, she developed severe truncal pruritus, and within 3 months, her external iliac stent graft had thrombosed. Allergy testing revealed nickel sensitivity. After medical therapy had failed, stent graft removal was performed, resulting in complete resolution of her symptoms. The present case demonstrates a rare allergic reaction to the nitinol in commercially available stent grafts. Pruritus and rash are rare reactions to stenting; however, a nitinol allergy should be considered for patients with no other identifiable primary source.
KW - Endovascular
KW - Nickel Allergy
KW - Nitinol
KW - Stent-graft
UR - http://www.scopus.com/inward/record.url?scp=85139724254&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85139724254&partnerID=8YFLogxK
U2 - 10.1016/j.jvscit.2022.08.013
DO - 10.1016/j.jvscit.2022.08.013
M3 - Article
C2 - 36248398
AN - SCOPUS:85139724254
SN - 2352-667X
VL - 8
SP - 562
EP - 564
JO - Journal of Vascular Surgery Cases, Innovations and Techniques
JF - Journal of Vascular Surgery Cases, Innovations and Techniques
IS - 4
ER -