Seraph 100 Microbind Affinity Blood Filter Does Not Clear Antibiotics: An Analysis of Antibiotic Concentration Data from PURIFY-OBS

Jesse P. DeLuca, Daniel J. Selig, Pooja Vir, Chau V. Vuong, Jeffrey Della-Volpe, Ian M. Rivera, Caroline Park, Benjamin Levi, Kathleen P. Pratt, Ian J. Stewart

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

Introduction: Novel hemoperfusion systems are emerging for the treatment of sepsis. These devices can directly remove pathogens, pathogen-associated molecular patterns, cytokines, and other inflammatory markers from circulation. However, significant safety concerns such as potential antibiotic clearance need to be addressed prior to these devices being used in large clinical studies. Methods: Prospective, observational study of 34 participants undergoing treatment with the Seraph 100® Microbind Affinity Blood Filter (Seraph 100) device at 6 participating sites in the USA. Patients were included for analysis if they had a record of receiving an antibiotic concurrent with Seraph 100 treatment. Patients were excluded if there was missing information for blood flow rate. Blood samples were drawn pre- and post-filter at 1 h and 4 h after treatment initiation. These average pre- and post-filter time-concentration observations were then used to estimate antibiotic clearance in L/h (CLSeraph) due to the Seraph 100 device. Results: Of the 34 participants in the study, 17 met inclusion and exclusion criteria for the antibiotic analysis. Data were obtained for 7 antibiotics (azithromycin, cefazolin, cefepime, ceftriaxone, linezolid, piperacillin, and vancomycin) and one beta-lactamase inhibitor. Mean CLSeraph for the antibiotics investigated ranged from −0.57 to 0.47 L/h. No antibiotic had a CLSeraph statistically significant from 0. Discussion/Conclusion: The Seraph 100 did not significantly clear any measured antibiotic in clinical samples. These data give further evidence to suggest that these therapies may be safely administered to critically ill patients and will not impact concentrations of administered antibiotics.

Original languageEnglish (US)
Pages (from-to)284-290
Number of pages7
JournalBlood Purification
Volume53
Issue number5
DOIs
StatePublished - Jan 13 2024

Keywords

  • Anti-infective agents
  • Antibiotics
  • Hemoperfusion
  • Pharmacokinetics

ASJC Scopus subject areas

  • Hematology
  • Nephrology

Fingerprint

Dive into the research topics of 'Seraph 100 Microbind Affinity Blood Filter Does Not Clear Antibiotics: An Analysis of Antibiotic Concentration Data from PURIFY-OBS'. Together they form a unique fingerprint.

Cite this