Polyvinyl alcohol foam particle sizes and concentrations injectable through microcatheters

John D. Barr, Thomas J. Lemley, Constantine N. Petrochko

Research output: Contribution to journalArticlepeer-review

26 Scopus citations

Abstract

PURPOSE: Injection of different size ranges and concentrations of polyvinyl alcohol (PVA) foam particles was performed to determine the optimal and maximum size ranges and concentrations that may be injected reliably through microcatheters used for vascular embolization. MATERIALS AND METHODS: Eight different microcatheters were tested. Six different size ranges of PVA foam particles (three each from two manufacturers) at three different concentrations were tested with each catheter. All PVA foam particles were suspended in dilute low-osmolarity contrast material. For each size range and concentration of PVA foam particles, sixty 1-mL injections were made into each catheter tested. Continuous pressure monitoring was employed during all injections. RESULTS: For each catheter, the authors determined recommended maximum size ranges of PVA foam particles that may be injected with reasonable pressure and minimal risk of inadvertent catheter occlusion. Among the four catheters used with 0.014-0.016-inch guide wires, the authors found that PVA foam particles as large as 1,000-1,500 μm could be injected without catheter occlusion. PVA foam particles as large as 710-1,000 μm could be injected through one catheter used with 0.010-0.013-inch guide wires. Particles as small as 300-500 μm occluded the other catheter used with the same size guide wires. Both of the flow-directed microcatheters tested would allow injection of PVA form particles as large as 355-500 μm. CONCLUSION: Given the recent proliferation of microcatheters, improved understanding of maximum PVA foam particle size ranges and concentrations that can be injected through them will help prevent inadvertent catheter occlusion during embolization. Because the authors' results are based on a simulated high- flow vascular lesion, they may not be applicable in all circumstances. Embolization of low-flow lesions, however, should not require use of the maximum possible size and concentration of PVA foam particles.

Original languageEnglish (US)
Pages (from-to)113-118
Number of pages6
JournalJournal of Vascular and Interventional Radiology
Volume9
Issue number1 I
DOIs
StatePublished - 1998

Keywords

  • Catheters and catheterization, technology
  • Embolization
  • Polyvinyl alcohol

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Polyvinyl alcohol foam particle sizes and concentrations injectable through microcatheters'. Together they form a unique fingerprint.

Cite this