Purpose: To further develop MultiNet GRAPPA, a neural-network-based reconstruction, for lower SNR proton MRSI (1H MRSI) data using adapted undersampling schemes and improved training sets. Methods: 1H FID-MRSI data and an anatomical image for GRAPPA reconstruction were acquired in two slices in the human brain (n = 6) at 7T. MRSI data were retrospectively undersampled for a 4×, 6×, and 7× acceleration rate. Signal-to-noise, relative error (RE) between accelerated and fully sampled metabolic maps, RMS of the lipid artifacts, and fitting reliability were compared across acceleration rates, to the fully sampled data, and with different kinds and amounts of training images. Results: Training with semi-synthetic images resulted in higher SNR and lower lipid RMS relative to training with acquired images from one or several subjects. SNR increased with the number of semi-synthetic training images and the 4× accelerated data retains ∼30% more SNR than other accelerated data. Spectra reconstructed with 20 semi-synthetic averages retained ∼100% more SNR and had ∼5% lower lipid RMS than those reconstructed with the center k-space points of one image as was originally proposed for very high SNR MRSI data and had higher fitting reliability. The metabolite RE was lowest when training with 20-semi-synthetic training images and highest when training with the center k-space points of one image. Conclusion: MultiNet GRAPPA is feasible with lower SNR 1H MRSI data if 20-semi-synthetic training images are used at a 4× acceleration rate. This acceleration rate provided the best trade-off between scan time and spectral SNR.

Original languageEnglish (US)
Pages (from-to)1500-1515
Number of pages16
JournalMagnetic resonance in medicine
Volume88
Issue number4
DOIs
StatePublished - Oct 2022

Keywords

  • GRAPPA
  • MRSI
  • acceleration
  • metabolite mapping
  • neural networks

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Improved signal-to-noise performance of MultiNet GRAPPA 1H FID MRSI reconstruction with semi-synthetic calibration data'. Together they form a unique fingerprint.

Cite this