T1 - Imaging performance evaluation of full and binning acquisition modes in a prototype CBCT system equipped with the TFT X-ray detector
AU - Seo, Chang Woo
AU - Cha, Bo Kyung
AU - Yang, Keedong
AU - Jeon, Seongchae
AU - Huh, Young
AU - Park, Justin C.
AU - Song, Bongyong
AU - Song, William Y.
AU - Lee, Byeonghun
N1 - Funding Information:
This research was supported by research project of Korea Electrotechnology Research Institute (KERI) funded by the Ministry of Knowledge Economy (No. 14-12-N0201-02 ).
Publisher Copyright:
© 2014 Elsevier B.V. All rights reserved.
PY - 2014/11/21
Y1 - 2014/11/21
N2 - The projection number, acquisition time, radiation dose, and full and 2×2 binning modes of the thin film transistor (TFT) X-ray imaging detector were evaluated for image quality in a prototype cone-beam computed tomography (CBCT) built for medical applications. The detector is an amorphous silicon (a-Si) based TFT X-ray detector (PaxScan 4030CB, Varian, Inc.). The a-Si based TFT X-ray detector has a 397×298 mm2 active area with 194 μm pixel pitch and 2048×1536 pixel, of which, 388 μm pixel pitch and 1024×768 pixel were used for the 2×2 binning mode. The Feldkamp, Davis, and Kress (FDK) reconstruction algorithm was used to generate 3D images, and the comparisons were made between different modes of acquisition. The American Association of Physicists in Medicine (AAPM) computed tomography (CT) performance phantom (model 610, CIRS) and the chest phantom (model 76-683, Nuclear Associates) were used to evaluate the image quality.
AB - The projection number, acquisition time, radiation dose, and full and 2×2 binning modes of the thin film transistor (TFT) X-ray imaging detector were evaluated for image quality in a prototype cone-beam computed tomography (CBCT) built for medical applications. The detector is an amorphous silicon (a-Si) based TFT X-ray detector (PaxScan 4030CB, Varian, Inc.). The a-Si based TFT X-ray detector has a 397×298 mm2 active area with 194 μm pixel pitch and 2048×1536 pixel, of which, 388 μm pixel pitch and 1024×768 pixel were used for the 2×2 binning mode. The Feldkamp, Davis, and Kress (FDK) reconstruction algorithm was used to generate 3D images, and the comparisons were made between different modes of acquisition. The American Association of Physicists in Medicine (AAPM) computed tomography (CT) performance phantom (model 610, CIRS) and the chest phantom (model 76-683, Nuclear Associates) were used to evaluate the image quality.
KW - Full and 2×2 binning modes
KW - Imaging performance
KW - Medical application
KW - Prototype CBCT system
KW - TFT X-ray detector
UR - http://www.scopus.com/inward/record.url?scp=84908347735&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84908347735&partnerID=8YFLogxK
U2 - 10.1016/j.nima.2014.05.024
DO - 10.1016/j.nima.2014.05.024
M3 - Article
AN - SCOPUS:84908347735
SN - 0168-9002
VL - 765
SP - 280
EP - 284
JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
ER -