Abstract
Background and Study Aims: Use of ultrathin (ut) endoscopes is sometimes limited by their design (outer diameter, tip angulation and image quality). New ut-endoscopes are being designed to address these limitations. A new utendoscope, XGIF-PV70N5 (Olympus America Inc.), with an outer diameter of 5.5 mm, 4-way angulation and narrow band imaging has recently been introduced. In this study, we report our subjective experience with this prototype ut-endoscope and discuss the practical uses of its new features. Patients and Methods: In this prospective case series, ut-endoscopy was performed on selected patients scheduled for endoscopy at a tertiary referral center. Results: Ut-endoscopy was successfully performed on 15 patients with varied indications. The 4-way angulation proved advantageous but not necessary. Narrow band imaging was not very useful in this group of patients. The image quality was satisfactory. Conclusions: The prototype ut-endoscope can be used for a broad range of indications. The 4-way angulation may prove advantageous in specific situations and the role of narrow-band imaging remains to be defined.
Original language | English (US) |
---|---|
Pages (from-to) | 909-914 |
Number of pages | 6 |
Journal | Digestive Diseases and Sciences |
Volume | 56 |
Issue number | 3 |
DOIs | |
State | Published - Mar 1 2011 |
Keywords
- Cholangioscopy
- Small caliber endoscope
- Ultrathin endoscopy
- Unsedated endoscopy