Abstract
The 3M Company recently produced a special version of the type 6702 seed for use in animal studies of ocular melanoma. The seed consists of a single 1–125 impregnated ion exchange resin sphere encapsulated in a 3.0-mm-long titanium cylinder, as opposed to the normal 4.5-mm cylinder containing from three to five resin spheres. Monte Carlo simulations and measurements show that the dose distribution around the 3-mm capsule differs in places by up to 10% from that of the 4.5-mm seed. In addition, the two seed types differ by 12% in the ratio of dose in tissue along the transverse seed axis to apparent activity. These differences must be taken into account when using these seeds.
Original language | English (US) |
---|---|
Pages (from-to) | 460-463 |
Number of pages | 4 |
Journal | Medical physics |
Volume | 17 |
Issue number | 3 |
DOIs | |
State | Published - May 1990 |
Keywords
- 1-125 dosimetry brachytherapy ocular melanoma Monte Carlo
ASJC Scopus subject areas
- Biophysics
- Radiology Nuclear Medicine and imaging