Cyclic variation of T1 in the myocardium at 3 T

Janaka Wansapura, William Gottliebson, Eric Crotty, Robert Fleck

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

Standard methods of longitudinal relaxation (T1) measurements in the heart produce only one T1 map of the myocardium, usually at the end diastole (ED). In this article, we investigated the feasibility of using a dual flip angle fast gradient echo technique in the steady state to generate a movie of T1 maps in the myocardium during the cardiac cycle. The effects of nonideal slice profile and transient steady state on the T1 measurements were evaluated by Bloch simulations. Based on these results, we introduce a linear correction to the measured T1 values, which was validated by phantom experiments. In vivo T1 cine maps in healthy volunteers show 70±7% drop in T1 from the ED to the end systole in the septum and a 43±13% drop in the left ventricular lateral wall. With further improvements, this technique could be used to assess the myocardial blood volume changes during the cardiac cycle.

Original languageEnglish (US)
Pages (from-to)889-893
Number of pages5
JournalMagnetic Resonance Imaging
Volume24
Issue number7
DOIs
StatePublished - Sep 2006
Externally publishedYes

Keywords

  • Cardiac MRI
  • Myocardium
  • T1

ASJC Scopus subject areas

  • Biophysics
  • Biomedical Engineering
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Cyclic variation of T1 in the myocardium at 3 T'. Together they form a unique fingerprint.

Cite this