Chronic monitoring of intracranial pressure using an in vivo calibrating sensor: Experience in patients with pseudotumor cerebri

P. R. Cooper, S. Moody, F. Sklar

Research output: Contribution to journalArticlepeer-review

12 Scopus citations

Abstract

A number of reliable methods are available for the short term monitoring of intracranial pressure (ICP). However, the danger of infection and the need for external connections make the use of these methods for extended periods inappropriate. Most previously described devices for chronic monitoring of ICP are of limited value because of uncertain drift from zero-point readings. The Hittman-Meyer ICP sensor approaches the ideal for a long term monitoring device. It is fully implantable and is accurate over a wide range of pressures and for extended periods. It is designed so that in vivo calibration may be performed in a noninvasive fashion. The authors report the use of this device in eight patients with pseudotumor cerebri. It has functioned accurately for periods ranging up to 14 months after implantation and has proven to be a valuable aid in the management of these patients.

Original languageEnglish (US)
Pages (from-to)666-670
Number of pages5
JournalNeurosurgery
Volume5
Issue number6
DOIs
StatePublished - 1979