TY - JOUR
T1 - Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging
T2 - A comparison study to conventional polychromatic 120-kVp images
AU - Rassouli, Negin
AU - Chalian, Hamid
AU - Rajiah, Prabhakar
AU - Dhanantwari, Amar
AU - Landeras, Luis
N1 - Funding Information:
This study was funded by institutional research grant from Philips Healthcare.
Publisher Copyright:
© 2017, Springer Science+Business Media New York.
PY - 2017/10/1
Y1 - 2017/10/1
N2 - Purpose: To evaluate the image quality of 70-keV virtual monoenergetic (monoE) abdominal CT images compared to 120-kVp polychromatic images generated from a spectral detector CT (SDCT) scanner. Methods: This prospective study included generation of a 120-kVp polychromatic dataset and a 70-keV virtual monoE dataset after a single contrast-enhanced CT acquisition on a SDCT scanner (Philips Healthcare) during portal venous phase. The attenuation values (HU), noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in the liver, spleen, pancreas, kidney, aorta, portal vein, and muscle. The subjective image quality including noise, soft tissue contrast, sharpness, and overall image quality were graded on a 5-point Likert scale by two radiologists independently (1-worst image quality, 5-best image quality). Statistical analysis was performed using paired sample t test and Fleiss’s Kappa. Results: Fifty-five patients (54.3 ± 16.8 y/o; 28 M, 27 F) were recruited. The noise of target organs was significantly lower in virtual monoE images in comparison to polychromatic images (p < 0.001). The SNR and CNR were significantly higher in virtual monoE images (p < 0.001 for both). Subjective image quality of 70-keV virtual monoE images was significantly better (p < 0.001) for all evaluated parameters. Median scores for all subjective parameters were 3.0 versus 4.0 for polychromatic vs virtual monoE images, respectively. The inter-reader agreement for overall image quality was good (Kappa were 0.767 and 0.762 for polychromatic and virtual monoE images, respectively). Conclusion: In abdominal imaging, 70-keV virtual monoE CT images demonstrated significantly better noise, SNR, CNR, and subjective score compared to conventional 120-kVp polychromatic images.
AB - Purpose: To evaluate the image quality of 70-keV virtual monoenergetic (monoE) abdominal CT images compared to 120-kVp polychromatic images generated from a spectral detector CT (SDCT) scanner. Methods: This prospective study included generation of a 120-kVp polychromatic dataset and a 70-keV virtual monoE dataset after a single contrast-enhanced CT acquisition on a SDCT scanner (Philips Healthcare) during portal venous phase. The attenuation values (HU), noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in the liver, spleen, pancreas, kidney, aorta, portal vein, and muscle. The subjective image quality including noise, soft tissue contrast, sharpness, and overall image quality were graded on a 5-point Likert scale by two radiologists independently (1-worst image quality, 5-best image quality). Statistical analysis was performed using paired sample t test and Fleiss’s Kappa. Results: Fifty-five patients (54.3 ± 16.8 y/o; 28 M, 27 F) were recruited. The noise of target organs was significantly lower in virtual monoE images in comparison to polychromatic images (p < 0.001). The SNR and CNR were significantly higher in virtual monoE images (p < 0.001 for both). Subjective image quality of 70-keV virtual monoE images was significantly better (p < 0.001) for all evaluated parameters. Median scores for all subjective parameters were 3.0 versus 4.0 for polychromatic vs virtual monoE images, respectively. The inter-reader agreement for overall image quality was good (Kappa were 0.767 and 0.762 for polychromatic and virtual monoE images, respectively). Conclusion: In abdominal imaging, 70-keV virtual monoE CT images demonstrated significantly better noise, SNR, CNR, and subjective score compared to conventional 120-kVp polychromatic images.
KW - Dual-layer CT
KW - Monoenergetic image
KW - SDCT
KW - Spectral detector CT
UR - http://www.scopus.com/inward/record.url?scp=85017615148&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85017615148&partnerID=8YFLogxK
U2 - 10.1007/s00261-017-1151-2
DO - 10.1007/s00261-017-1151-2
M3 - Article
C2 - 28421243
AN - SCOPUS:85017615148
SN - 2366-004X
VL - 42
SP - 2579
EP - 2586
JO - Abdominal Radiology
JF - Abdominal Radiology
IS - 10
ER -