TY - JOUR
T1 - An algorithm for estimation and correction of anisotropic magnification distortion of cryo-EM images without need of pre-calibration
AU - Yu, Guimei
AU - Li, Kunpeng
AU - Liu, Yue
AU - Chen, Zhenguo
AU - Wang, Zhiqing
AU - Yan, Rui
AU - Klose, Thomas
AU - Tang, Liang
AU - Jiang, Wen
N1 - Funding Information:
This work was supported in part by NIH grant (1R01AI111095). Cryo-EM images were taken at the Purdue Cryo-EM Facility ( http://cryoem.bio.purdue.edu ).
Publisher Copyright:
© 2016 Elsevier Inc.
PY - 2016/8/1
Y1 - 2016/8/1
N2 - Anisotropic magnification distortion of TEM images (mainly the elliptic distortion) has been recently found as a potential resolution-limiting factor in single particle 3-D reconstruction. Elliptic distortions of ∼1–3% have been reported for multiple microscopes under low magnification settings (e.g., 18,000×), which significantly limited the achievable resolution of single particle 3-D reconstruction, especially for large particles. Here we report a generic algorithm that formulates the distortion correction problem as a generalized 2-D alignment task and estimates the distortion parameters directly from the particle images. Unlike the present pre-calibration methods, our computational method is applicable to all datasets collected at a broad range of magnifications using any microscope without need of additional experimental measurements. Moreover, the per-micrograph and/or per-particle level elliptic distortion estimation in our method could resolve potential distortion variations within a cryo-EM dataset, and further improve the 3-D reconstructions relative to constant-value correction by the pre-calibration methods. With successful applications to multiple datasets and cross-validation with the pre-calibration method, we have demonstrated the validity and robustness of our algorithm in estimating the distortion; correction of the elliptic distortion significantly improved the achievable resolutions by ∼1–3 folds and enabled 3-D reconstructions of multiple viral structures at 2.4–2.6 Å resolutions. The resolution limits with elliptic distortion and the amounts of resolution improvements with distortion correction were found to strongly correlate with the product of the particle size and the amount of distortion, which can help assess if elliptic distortion is a major resolution limiting factor for single particle cryo-EM projects.
AB - Anisotropic magnification distortion of TEM images (mainly the elliptic distortion) has been recently found as a potential resolution-limiting factor in single particle 3-D reconstruction. Elliptic distortions of ∼1–3% have been reported for multiple microscopes under low magnification settings (e.g., 18,000×), which significantly limited the achievable resolution of single particle 3-D reconstruction, especially for large particles. Here we report a generic algorithm that formulates the distortion correction problem as a generalized 2-D alignment task and estimates the distortion parameters directly from the particle images. Unlike the present pre-calibration methods, our computational method is applicable to all datasets collected at a broad range of magnifications using any microscope without need of additional experimental measurements. Moreover, the per-micrograph and/or per-particle level elliptic distortion estimation in our method could resolve potential distortion variations within a cryo-EM dataset, and further improve the 3-D reconstructions relative to constant-value correction by the pre-calibration methods. With successful applications to multiple datasets and cross-validation with the pre-calibration method, we have demonstrated the validity and robustness of our algorithm in estimating the distortion; correction of the elliptic distortion significantly improved the achievable resolutions by ∼1–3 folds and enabled 3-D reconstructions of multiple viral structures at 2.4–2.6 Å resolutions. The resolution limits with elliptic distortion and the amounts of resolution improvements with distortion correction were found to strongly correlate with the product of the particle size and the amount of distortion, which can help assess if elliptic distortion is a major resolution limiting factor for single particle cryo-EM projects.
KW - Anisotropic magnification distortion
KW - Elliptic distortion
KW - Generalized 2-D alignment
KW - Single particle 3-D reconstruction
KW - jspr
UR - http://www.scopus.com/inward/record.url?scp=84992045306&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84992045306&partnerID=8YFLogxK
U2 - 10.1016/j.jsb.2016.06.003
DO - 10.1016/j.jsb.2016.06.003
M3 - Article
C2 - 27270241
AN - SCOPUS:84992045306
SN - 1047-8477
VL - 195
SP - 207
EP - 215
JO - Journal of Structural Biology
JF - Journal of Structural Biology
IS - 2
ER -