A quality assurance protocol for diffusion tensor imaging using the head phantom from American College of Radiology

Zhiyue J. Wang, Youngseob Seo, Jonathan M. Chia, Nancy K. Rollins

Research output: Contribution to journalArticlepeer-review

31 Scopus citations

Abstract

Purpose: To propose a quality assurance procedure for routine clinical diffusion tensor imaging (DTI) using the widely available American College of Radiology (ACR) head phantom. Methods: Analysis was performed on the data acquired at 1.5 and 3.0 T on whole body clinical MRI scanners using the ACR phantom and included the following: (1) the signal-to-noise ratio (SNR) at the center and periphery of the phantom, (2) image distortion by EPI readout relative to spin echo imaging, (3) distortion of high-b images relative to the b = 0 image caused by diffusion encoding, and (4) determination of fractional anisotropy (FA) and mean diffusivity (MD) measured with region-of-interest (ROI) and pixel-based approaches. Reproducibility of the measurements was assessed by five repetitions of data acquisition on each scanner. Results: The SNR at the phantom center was approximately half of that near the periphery at both 1.5 and 3 T. The image distortion by the EPI readout was up to 7 mm at 1.5 T and 10 mm at 3 T. The typical distortion caused by eddy currents from diffusion encoding was on the order of 0.5 mm. The difference between ROI-based and pixel-based MD quantification was 1.4% at 1.5 T and 0.3% at 3 T. The ROI-based MD values were in close agreement (within 2%) with the reference values. The ROI-based FA values were approximately a factor of 10 smaller than pixel-based values and less than 0.01. The measurement reproducibility was sufficient for quality assurance (QA) purposes. Conclusions: This QA approach is simple to perform and evaluates key aspects of the scanner performance for DTI data acquisition using a widely available phantom.

Original languageEnglish (US)
Pages (from-to)4415-4421
Number of pages7
JournalMedical physics
Volume38
Issue number7
DOIs
StatePublished - Jul 2011

Keywords

  • diffusion tensor imaging
  • phantom
  • quality assurance
  • quality control

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'A quality assurance protocol for diffusion tensor imaging using the head phantom from American College of Radiology'. Together they form a unique fingerprint.

Cite this